Procedure for the Calibration of Permanent Magnet Moving Coil (PMMC) Voltmeter:-

Calibration of Voltmeter Circuit
  1. Give connections as per the circuit diagram in above figure.

  2. Vary the voltage across the voltmeter so that it reads a rounded position, say 5 volts.

  3. Measure the voltage across the voltmeter by potentiometer and tally the voltmeter indication and potentiometer reading.

  4. Draw the calibration curve which is drawn between % error voltage Vs measured potentiometer readings. The expected curve is as follows:
    Calibration of Voltmeter using Potentiometer Expected Curve

  5. Repeat the above steps for different voltages and tabulate the results. The expected tabular readings are given below:

    S.NoVoltmeter ReadingPot. Reading m.f(v)Measured Pot. Reading = [(m.f)x 30/1.5]% error
    1.50.2494.980.4
    2. 100.4879.742.6
    3.150.736 14.721.86
    4.200.98719.74 1.3
    5.251.219524.392.44
    6.301.46129.222.6

Precautions

  • Connections should always be made with due care of polarity.

  • For measuring higher voltages better use a volt ratio box.

  • Standardize the potentiometer at regular intervals during prolonged tests to compensate it from any slight drift in the battery voltage.